FUNCTIONAL DESCRIPTION
IV-2
1. SYSTEM
NOTE:
Any reference to PD 1400 includes PD 1400, D 1400,
PD 2000, D 2000, and ZOLL 1600 unless otherwise
specified
Product Overview
The ZOLL Base PowerCharger
4x4
(with QuickCharge or AutoTest) is a battery charger
and testing system designed for management of PD 4410 battery packs used in ZOLL
Medical Corporation resuscitation devices.
The ZOLL Base PowerCharger
4x4
(with QuickCharge or AutoTest) provides four battery
charging/testing compartments. Up to four battery packs may be charged or tested in
any combination at one time.
The ZOLL Base PowerCharger
4x4
can be configured in either the
AutoTest
or
QuickCharge
mode. This configuration is performed by ZOLL Medical Corporation prior
to device shipment. The Base PowerCharger
4x4
is clearly labeled
AutoTest
or
QuickCharge
on the charger label directly beneath the product name. If the device is not
labeled, the QuickCharge instructions should be followed for proper device operation.
For questions concerning the chargers configuration please consult a ZOLL technical
service representative.
AutoTest
The ZOLL Base PowerCharger
4x4
with AutoTest automatically tests battery capacity with
each battery recharge. In addition, the ZOLL Base PowerCharger
4x4
with AutoTest
illuminates the
Batt. Ready
indicator when the battery is fully charged and capable of
powering a PD 1400/D 1400 or PD 2000/D 2000 for approximately 2.0 hours in Monitor
mode*. Fully charged batteries whose capacity is insufficient to power the devices for this
period of time will cause the
FAULT
light to illuminate. With AutoTest, the full charging
cycle is complete in eight (8) hours or less. The AutoTest feature virtually eliminates the
need to periodically test battery capacity since this test is performed each time a battery
is recharged.
QuickCharge
The ZOLL Base PowerCharger
4x4
with QuickCharge illuminates the Batt. Ready indicator
when the battery is fully charged. The full charging cycle is complete in four (4) hours or
less. The QuickCharge version provides
no
information regarding battery capacity
because it does
not
test the battery capacity with each recharge. The ZOLL Base
PowerCharger
4x4
with QuickCharge relies on the user to manually initiate the battery test
cycle at recommended intervals (see Manual Battery Testing Procedure) in order to
confirm adequate battery capacity.
Note:
Actual battery run time will vary from model to model depending upon the
amount of current that is drawn by each device. Refer to the Operator's Manual
of your model to determine the actual battery run time. For example, a battery
contains at least 1.5 ampere hours, while a ZOLL Defibrillator may draw
approximately .8 ampere of current. The run time of the battery is
calculated by dividing the battery capacity by the current drawn. In this example,
there is a minimum of 1.875 hours of battery run time for this device.
Summary of Contents for Base Powercharger 4x4
Page 1: ...9650 0072 01 Rev E ZOLL Base Powercharger 4x4 Service Manual...
Page 8: ...GENERAL INFORMATION I 7 SEE FILE PG7_C PM5 FOR THIS PAGE AND ALL OF THE SYMBOLS...
Page 11: ...GENERAL INFORMATION I 10...
Page 13: ...CHECKOUT PROCEDURES 9650 0072 II 2 Rev D...
Page 39: ...SCHEMATIC DRAWINGS V 2...
Page 49: ...COMPONENT LAYOUT DRAWINGS VI 2...
Page 53: ...DISASSEMBLY PROCEDURES VII 2...
Page 55: ...DISASSEMBLY PROCEDURES VII 4...
Page 57: ...DISASSEMBLY PROCEDURES VII 6...
Page 58: ...DISASSEMBLY PROCEDURES VII 7...
Page 60: ...DISASSEMBLY PROCEDURES VII 9...
Page 62: ...DISASSEMBLY PROCEDURES VII 11...
Page 63: ...DISASSEMBLY PROCEDURES VII 12...
Page 65: ...DISASSEMBLY PROCEDURES VII 14...
Page 67: ...DISASSEMBLY PROCEDURES VII 16...
Page 71: ...REPLACEMENT PARTS VIII 4...