www.zinexts.com MagPurix 24 EVO User Manual
11
/
35
WARNING
:
Use of other cables and accessories other than provided with the instrument
may negatively affect EMC performance and results in non-compliance.
3.2
Operating Environment / Conditions
Use the MagPurix 24 EVO System in a location that meets the following conditions:
The MagPurix 24 EVO System must be placed at a minimum distance of 5 – 10 cm from all
sidewalls and other instruments, to allow appropriate air circulation.
A location where power supply is provided with safety.
A location where the temperature ranges between 15 – 30°C, and humidity ranges from 30%
to 80% RH (non-condensing).
A location that is flat and stable, with no vibration.
A location away from direct sunlight. Block sunlight by closing curtains or blinds, if necessary.
A location which is well-ventilated and not dusty.
A location where the temperature does not go up and down suddenly. Warming a cold room
suddenly or moving the MagPurix 24 EVO System from a room with low temperature to a warm
room may cause condensation inside the device, resulting in abnormal operation.
A location where the temperature and humidity are kept within the specified range (far away
from water taps, water heaters, humidifiers, air-conditioners, and heaters).
A location far away from objects that generate strong magnetic fields (motors, transformers,
TV, audio speakers, magnets, etc.). Bringing the MagPurix 24 EVO System close to any type of
magnetic field may cause malfunction.
WARNING
:
Do not use the MagPurix 24 EVO System in a location where it is wet or can be splashed
with water. It may cause device failure, fire, or electric shock.
When relocating the MagPurix 24 EVO System, disconnect the plug from the socket first.
If the power cable is damaged, it may cause device failure, fire, injury, or electric shock.
Caution:
Do not use the MagPurix 24 EVO System in an unstable place such as a slanted surface
or a place subject to vibrations. It may cause injury or device failure.