www.zinexts.com MagPurix 24 EVO User Manual
21
/
35
5.3
Perform Extraction
Note:
Before operating the MagPurix 24 EVO System, please prepare all items necessary to run the
instrument, as described in Chapter 5.1.
◼
After logging in to the MagPurix 24 EVO Software, the Main Screen opens. Click the “Perform
Run” button, and on the “Perform Run” Screen click the “Extraction” button.
◼
Place reaction cartridges and all plastic consumables onto the worktable:
Important:
Place the samples and reagent cartridges (1 – 12 samples) starting from left to right
on the sample tube rack and the extraction module. Ensure that the reagent cartridges are
inserted tightly into the cartridge holder.