www.zinexts.com MagPurix 24 EVO User Manual
26
/
35
Schedule
Maintenance Type
Performed by
6-months to
12-months
interval
Preventive (for details of the procedures please refer to the
MagPurix 24 EVO
Service Manual)
Grease ball screws if required
Check motor function & axis calibration
Check air-tight status of the piston units
Cleaning of piston
UV light change annually
Service Engineer
6.1
Cleaning of Rubbish Tray and M Motor
Elevate M motor before the start of cleaning
◼
Clean the rubbish tray and M motor with mild detergent using a soft cloth. Then rinse and wipe
with deionized water using a soft cloth. Allow the parts to dry before use.
◼
Clean and disinfect the platform surface by wiping with deionized water followed by 75%
ethanol.
◼
Click “Finish” after cleaning to bring M motor back in position.