Error messages, troubleshooting,
disposal
Error messages / Troubleshooting
15
Page 32
Error codes
The PremoPort
Four
has various error codes which quickly provide information
about the cause of the error. The error is displayed on the device, saved, and
transferred with the measurement data to Diagnostik
Suite
.
Error
-
code
Description
Possible cause
01
Measurement
discontinued
Due to an arm movement or a vibration, the
device needed more than the maximum
possible measurement time. For the safety of
the patient, the device discontinues a
measurement in progress after 120 sec. if the
measurement is not being performed
successfully.
02
Manually discontinued
The patient pressed a button on the device to
discontinue the measurement.
03
Batteries dead
The batteries died during the measurement.
The remaining charge was not sufficient to
continue the measurement.
04
Batteries replaced
The batteries were replaced during the
ambulatory measurement.
09
External interference
External electrical interferences or static
discharges caused an interruption of the
measurement.
31
Cuff not connected
The cuff was not connected to the device when
the blood pressure measurement was started.
32
Airway problem
There was a problem in the air tube or in the
pneumatic system. In general, this means the
air tube was blocked or pinched.
33
Cuff not airtight
The system was not able to maintain air
pressure. In general, this means the cuff is not
airtight.
34
Cuff not on the arm
The cuff was not on the arm or it was placed
too loosely.
90, 99 Device error
Internal device error; please notify the service
personnel.