Functional test
13
Page 29
13.
Functional test
Prior to initial start-up, perform a functional test to check the functionality of the
ambulatory blood pressure system.
Functional test
PremoPort
Four
Proceed as follows:
Connect the individual components as described in the previous chapters.
Place the disposable or rechargeable batteries in the battery compartment
and close the lid. The voltage of the disposable or rechargeable batteries
will be shown on the display for approx. 5 seconds. (e.g. 2_23 corresponds
to 2.23 V). Then the display changes to show the time.
To check the display and blood pressure measurement unit, place the
blood pressure cuff on the upper arm and connect the cuff to
PremoPort
Four
. Briefly push the Start/Stop button. All segments of the
LCD display can be seen for approx. 3 seconds.
Then the pump process begins. The increasing numerical values on the
display and an increase in pressure in the cuff can be seen.
After a successful measurement, the systolic blood pressure, the diastolic
blood pressure and the pulse are displayed sequentially six times
consecutively.
Disconnect the examination device(s) and remove the
disposable or rechargeable batteries.
Note:
In the event of malfunction or visible damage, do not use the ambulatory blood
pressure measurement system. Contact the service department of
Zimmer MedizinSysteme.