The device is not listed in attachment 1 but it is listed in attachment 2 of the
Medical Device Operator Ordinance (MPBetreibV). According to section 14
MPBetreibV, the device must undergo a metrological control (MTK) every two
years. The result must be documented in the medical device book and identified
by a test seal on the device indicating the year during which the next control is
due. The metrological control may only be performed by an approved testing
facility. Contact Zimmer MedizinSysteme in this regard.
In Germany, the regulation regarding electrical systems and equipment
(German Social Accident Insurance (DGUV) Regulation 3), as amended, must
additionally be observed.
Note:
This information applies to the operation of the unit in Germany. Please
consider divergent national regulations in your country, if applicable.