Manufacturer’s EMC Declaration
18
Page 43
The main features of the enPulsPro are: the trouble-free delivery of shockwaves and operation of all functions. Uninterrupted
operation is not required for the intended application.
Guidelines and manufacturer’s declaration – electromagnetic interference resistance
The enPulsPro is intended for use in one of the electromagnetic environments specified below. The customer or user of the
enPulsPro should ensure that it is only used in such an environment.
Interference
tests
IEC 60601 test level Conformity
level
Electromagnetic environment – guidelines
Conducted
RF
interference
in
accordance
with IEC
61000-4-6
Radiated RF
disturbance
according to
IEC 61000-
4-3
3 V
RMS
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
RMS
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile RF communications
equipment (including cables) should be used no
closer to the
enPulsPro than the recommended separation
distance, which is calculated according to the
equation for the transmission frequency.
Recommended separation distance:
d = 1.2 √P
d = 0.35 √P for 80–800 MHz
d = 0.7 √P for 800 MHz to 2.5 GHz
where P is the nominal rating of the transmitter in
watts (W) in accordance with the manufacturer of
the transmitter and d is the recommended
separation distance in metres (m).
In accordance with an on-site investigation
a
, for all
frequencies, the field strength of a stationary radio
transmitter should be less than the coincidence
level
b
.
In the vicinity of devices marked with the following
symbol, errors may occur:
NOTE 1: At 80 and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all cases. The spread of electromagnetic variables is affected by the absorption
and reflection of structures, objects, and people.