Operating Instructions
8.3 Performing the Treatment
8
Page 12
Note:
All buttons, menus, and sub-menus can be activated directly on the screen by
finger pressure.
Program start-up
Pressing the "Therapy" button on the start screen opens the therapy screen.
Select applicator
Select an applicator to suit your desired treatment and screw it correctly into
the handpiece.
Attach
handpiece/applicator
Place the handpiece on the selected treatment point/area. To prevent friction
on the skin, the enPulse Lotion can be applied to the treatment area before
treatment.
Caution
!
When using lubricants, the applicator must be covered with the protective
silicone cap.
Set pulse energy
Set the pulse energy using the left controller.
Note:
The enPulsPro offers two methods of pulse emission.
Pulse emission with pre-set number of pulses
For pulse emission with pre-set number of pulses, the treatment is terminated
by the device after the pre-set number of pulses has been emitted.
The footswitch is deactivated and the emission of pulses is no longer possible.
Treatment can be continued by resetting the current pulse rate or by adjusting
the pre-set.
Pulse emission without pre-setting the number of pulses
With pulse output without pre-setting the number of pulses, the treatment is not
terminated by the device. Pulses are emitted as long as the footswitch is
activated.
For pulse emission without pre-set, only the ascending counting direction is
active.
Start of treatment
The treatment starts once the footswitch has been activated.
Note:
The shockwave should be activated via the footswitch only after placing the
handpiece on the patient.
End of treatment
By de-activating the footswitch, the treatment is interrupted or terminated.
Note:
During the treatment, the patient must be monitored carefully. If any issues
arise, the treatment should be adjusted or terminated.