Operating Instructions
8.4 Display and Buttons
8
Page 13
Description of the
display elements
and buttons
(1) Pulse energy
Displays the set pulse energy. In the case of active treatment, the bar graph is
filled in. The pulse energy can be set before and during the emission of pulses.
The pulse energy can be adjusted from 60 to 185 mJ in 10 mJ increments.
(2) Selection of
handpiece
Connecting one handpiece:
If only one handpiece is connected, the channel of the connected handpiece is
shown in the selection window. The handpiece is activated automatically
regardless of the connected channel.
Connection of two handpieces:
If two handpieces are connected, the handpiece on Channel I is initially
activated.
The desired handpiece can be activated by directly selecting the handpiece in
the selection window. The activated handpiece is highlighted.
(3) VAS Pre
Activating the VAS Pre window opens the screen to measure the subjective
sensation of pain prior to the treatment.
(4) VAS Post
Activating the VAS Post window opens the screen to measure the subjective
sensation of pain following the treatment.
The precise procedure for this measurement is described in detail in Chapter
8.8.