Operating Instructions
8.1 Device Description
8
Page 10
Handpiece
temperature
monitoring /
regulation
The generation of mechanical impact energy generates considerable heat
within the handpiece. A temperature switch has been integrated to prevent a
reduction in the life of the handpiece. In the case of overheating, this forces
the cooling of the handpiece through an internal shutdown.
In addition to temperature monitoring, the enPulsPro regulates the
temperature using a sensor in the handpiece. The fan in the handpiece is
started when activated by the footswitch and automatically stops when it
reaches a certain temperature.
When the temperature reaches a critical limit, a cooling phase is initiated. This
is indicated in the display by the following message:
"Overheating of the applicator. Please allow the applicator to cool down. "
An output pulse is no longer possible.
After confirming the message with "OK", the treatment screen comes to the
foreground. In the selection window of the handpiece, the handpiece is shown
a deactivated. The time required for the applicator to cool down is displayed.
Applicator heads
Three different applicator heads are available for the treatment.
Changing the
applicator heads
To change the applicator head, hold the handpiece with one hand and turn the
applicator head anti-clockwise with the other hand. Turn the desired head
clockwise until it stops.
Note
:
The applicator heads are wear parts and must be replaced after a certain
operating life (see Chapter 14.2 Maintenance).
Footswitch
Place the footswitch in such a way that it can be easily reached during the
treatment. The operating element of the switch is independent of direction; a
precise alignment of the footswitch is therefore unnecessary.
To avoid damage, ensure that only slight pressure is exerted on the switch.
The operation is done with the forefoot and not the heel.
The switch does not have any lock i.e. the activation remains only as long as
pressure is applied to the switch.