Documentation set
12 Optional accessories
SL 800
12.9 Tube adapter 20°
000000-2268-355-DokS-GB-300719
79 / 94
12.9 Tube adapter 20°
The tube adapter 20° causes the convergent or parallel tube to
slant upwards by 20°, relieving strain on the neck when working.
Assembly:
Action
1. Switch off the device at the power supply unit.
2. Lock the instrument base using the manual brake.
3. Hold the binocular tube while loosening the threaded pin at the
top of the convenience interface with the size 1.5 screwdriver,
supplied.
4. Tilt the binocular tube downwards out of the convenience
interface.
5. Put the binocular tube aside.
6. Insert the tube adapter 20° into the convenience interface and
tilt it upwards.
Summary of Contents for SL 800