Documentation set
4 Installation
SL 800
4.2 Installing the device
000000-2268-355-DokS-GB-300719
29 / 94
delivery package). Ensure that the deflectors do not protrude
beyond the table top and that the cable of the fixation lamp is
not pinched.
Figure 10: Position the deflectors laterally to the head rest feet on the
patient's side of the examination unit.
10. If applicable, remove additional system components and attach
them according to the instructions in the Optional accessories
[
65] chapter.
11. If the device is mounted on an instrument table qualified by
ZEISS, attach the power supply unit to the intended position
underneath the table top. Use the four screws provided.
12. When using the device on an examination unit, connection
cables can be extended using the optional cable extension set.
The technician of the manufacturer of the examination unit
should perform the installation.
13. If necessary, change the factory settings of the device on the
power supply unit to the switch allocation described in the
Configuration switch on power supply unit [
25] chapter by
changing the respective configuration switch settings using an
object with a rounded tip.
Summary of Contents for SL 800
Page 1: ...SL 800 Slit lamp Documentation set ...
Page 6: ...Empty page for your notes ...
Page 32: ...Empty page for your notes ...
Page 34: ...Empty page for your notes ...
Page 36: ...Empty page for your notes ...
Page 44: ...Empty page for your notes ...
Page 52: ...Empty page for your notes ...
Page 64: ...Empty page for your notes ...
Page 88: ...Empty page for your notes ...
Page 90: ...Empty page for your notes ...
Page 92: ...Empty page for your notes ...