Documentation set
3 Description of the device
SL 800
3.5 Control elements
000000-2268-355-DokS-GB-300719
21 / 94
Pos. Name
Explanation
7
Button to release
decentration of slit
image
Allows decoupling of observation and
illumination unit when released. When
fixed, the observation and illumination unit
is coupled, i.e. centered.
8
Lever for vertical
inclination angle
Allows setting of the angle of incidence
from 0° to 20° in steps of 5°
9
Rotary knob for slit
width adjustment
Setting range: continuously from 0 mm to
12 mm, with scale from 0 mm to 0.2 mm
10
Scale for angular
difference
Scale indicating angular difference between
illumination and observation with click
stops at -60°/-45°/0°/+45°/+60°
11
Locking lever
Lever to fixate the angular difference
between illumination and observation