CHAPTER 2 - SETUP REQUIREMENTS
ZEISS
Environmental Requirements
LSM 880
46
000000-2071-466
11/2014 V_05
5
Environmental Requirements
For operation the system has to be placed in a closed room.
1. Operation, specified performance
T = 22 °C ± 3 °C without interruption (24 h a day
independently whether system is operated or
switched-off)
It has to be ensured that the air-flow of the
air-conditioning is not directed at the system.
2. Operation, reduced performance
T = 15 °C to 35 °C, any conditions different from
item 1. and 5.
3. Storage, less than 16 h
T = -20 °C to 55 °C
4. Storage, less than 6 h
T = -20 °C to 55 °C
5. Temperature gradient
± 0.5 °C/h
6. Warm up time
1 h, for high-precision and/or long-term measure-
ments
≥
3 h
7. Relative humidity
< 65 % at 30 °C
8. Operation altitude
max. 2000 m
9. Loss of heat (without Ti:Sa or OPO laser)
2 kW
10. Vibrations under operation conditions
(with system table)
5 µm pp at 5 Hz
10 µm pp at 10 Hz
10 µm pp at 20 Hz
11. Shipping shock (LSM 880 box)
10 g