11. Switching off the system
11-1.
Turn off the Lasers
●
DPSS(561nm), HeNe (543nm, 594nm, 633nm), MaiTai(Spectra physics)
:
Open the
Setup Manager - Laser
tool.
Turn off
the lasers from pull down menu.
●
Chameleon (Coherent)
:
Open the
Setup Manager - Laser
tool.
Turn off
the laser from pull down menu.
After that, turn the
key switch
from On to
Standby.
●
Argon (458/488/514nm)
:
Switch off the Ar laser with first the control knob (Fig.2/
④
), second the idle-run-
switch (Fig. 2/
③
) and third the key switch (Fig. 2/
②
), then
wait until the fan of the
Argon laser has switched off (about 5min)
. Don’t switch off the main power yet.
After cooling finished,
switch off the Ar laser
by the main power switch on the
power supply (Fig. 2/
①
).
11-2.
Switch off the
X-Cite 120 lamp
or the
HBO 100
mercury burner.
11-3.
Clean the Objective lens if you used immersion medium (Oil, ImmersolW)
. And
set the position of the objective lens to the lowest magnification for the next user.
11-4.
Click on the
File
-
Exit
button to leave the ZEN software.
11-5.
Shut down the computer.
11-6.
Turn off the Power remote switch.
①
Turn off the
Components
switch and the
System/PC
switch.
②
Turn off the
Main Switch.