START-UP 

Colibri.2 

Construction of the illumination system 

Carl Zeiss 

 
 

 
 

07/2014 

423052-7244-001 

19 

3.4

 

Construction of the illumination system 

3.4.1

 

Colibri.2 lamp module  

The  Colibri.2  lamp module can be equipped with 
up to four LED modules (Fig.  3-5/

1

) and three 

beam combiners (Fig. 3-5/

2

). 

The LED modules and beam combiners can be 

switched by the customer. 
When using an additional fiber coupled white light 

source, it is possible to toggle between the lamp 

module and the external white light source by 
means of a motorized built-in switching prism  
(Fig. 3-5/

3

, covered). 

 
 
 
 
 

3.4.2

 

Control unit 

The control unit of the lamp module contains the 

control and the power electronics.

 

At the front of the device there is the ON/OFF 
switch 

Power

 (Fig. 3-6/

1

 with pilot lamp), as well 

as a 15-pin subD port (Fig.  3-6/

2

) for the control 

panel. 
 
The rear of the control unit contains: 

 

several BNC ports; the trigger cable leading 
to the white light source has to be inserted 
in 

OUT 1

 (Fig. 3-7/

1

 

the power connection socket (Fig. 3-7/

2

 

two CAN ports (Fig.  3-7/

3

) for the 

connection to the microscope 

 

the 25-pin subD port (Fig.  3-7/

4

) for 

connecting the Colibri.2 lamp module 

 

the USB port (Fig. 3-7/

5

) for the connection 

to a PC. 

 
 
 

 

Fig. 3-5 

Construction of the lamp module 

 

Fig. 3-6 

Control unit, front  

 
 

 

Fig. 3-7 

Control unit, rear  

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for Colibri.2

Page 1: ...s underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos...

Page 2: ...Illumination System Colibri 2 Operating Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 3: ...rement the violating party will be held liable for damage All rights reserved in the event of a patent approval or utility model registration All company and product names mentioned in this operating...

Page 4: ...4 1 Switching on the control unit 22 4 2 Functions of the control panel 22 4 3 Functions of the white light source HXP 120 V 23 5 MODIFYING THE LAMP MODULE 24 5 1 Removing the lamp module from the mi...

Page 5: ...V rear 21 Fig 4 1 Control unit 22 Fig 4 2 Control panel with lamp module being controlled 22 Fig 4 3 Various indications on the control panel display when using the Colibri 2 illumination system and t...

Page 6: ...nation adapter may only be removed by Carl Zeiss Service not a customer interface Otherwise there is a risk of eye damage upon laser coupling WARNING of possible laser radiation The safety ring on the...

Page 7: ...ination system in connection with the microscope please pay attention to the operating manual of the microscope Do not operate the Colibri 2 illumination system in explosion prone areas Keep combustib...

Page 8: ...vering ventilation slats of the control unit may lead to heat build up that may damage the device and in extreme cases cause a fire Always keep the ventilation slats clear and ensure that no objects e...

Page 9: ...he manufacturer of the white light source Otherwise there is danger of burns and explosion when changing the bulb Defective devices must not be disposed of with the household waste Dispose them of in...

Page 10: ...abels Carl Zeiss 07 2014 423052 7244 001 9 1 2 Warning labels Standard labeling Fig 1 1 Warning labels on the Colibri 2 illumination system Artisan Technology Group Quality Instrumentation Guaranteed...

Page 11: ...y screws Fig 2 1 1 and 2 on the lamp module and Colibri 2 illumination adapter may only be removed by Carl Zeiss Service not a customer interface Otherwise there is a risk of eye damage upon laser cou...

Page 12: ...rticular and improper use The manufacturer of the device is not liable for damage caused by faulty operation negligence or any other tampering with the device in particular as a result of removing or...

Page 13: ...les with different emission wavelengths can be used individually or combined in the lamp module LED modules are exchangeable at present ten different LED modules are available No alignment required LE...

Page 14: ...ts overview Carl Zeiss 07 2014 423052 7244 001 13 2 2 Components overview Fig 2 1 Components overview of the Colibri 2 illumination system Artisan Technology Group Quality Instrumentation Guaranteed 8...

Page 15: ...ation max 75 at 35 C Altitude of installation site max 2000 m Atmospheric pressure 800 hPa to 1060 hPa Pollution degree 2 Operating data Operating environment closed room Protection class I Protection...

Page 16: ...es are supplied separately in customary packaging Remove all parts from the packaging and check them for completeness according to the delivery note For use with confocal Laser Scanning Microscopes th...

Page 17: ...n subD port of the control unit see Fig 3 6 2 Connect the control unit see Fig 3 7 2 with the power cable to a power outlet An external white light source e g HXP 120 V see Fig 3 10 can be connected a...

Page 18: ...s 07 2014 423052 7244 001 17 3 3 Connection diagram 3 3 1 Connection diagram for stand alone operation Fig 3 3 Connection diagram for stand alone operation Artisan Technology Group Quality Instrumenta...

Page 19: ...iagram with PC and motorized microscope for Fast Acquisition Colibri 2 can be connected alternatively via USB or via CAN Fig 3 4 Connection diagram with PC and motorized microscope for Fast Acquisitio...

Page 20: ...he control unit of the lamp module contains the control and the power electronics At the front of the device there is the ON OFF switch Power Fig 3 6 1 with pilot lamp as well as a 15 pin subD port Fi...

Page 21: ...to control the LED modules switch over to the external white light source and control the shutter of the white light source HXP 120 V The control panel has to be connected to the 15 pin subD port see...

Page 22: ...2500 hours Change lamp Flashing red Lamp defective At the rear there are the power connection socket Fig 3 11 2 and the fiber optics outlet Fig 3 11 3 Besides that the trigger cable port Trigger in Fi...

Page 23: ...us ON and for OFF is indicated The LED modules are individually switched on off by pressing the control elements 1 to 4 Fig 4 2 5 down These control elements are also used to regulate the intensity by...

Page 24: ...ior to switching on the white light source please make sure that the liquid light guide has been properly positioned and fixed to the Colibri 2 lamp module Switch on or off the external white light so...

Page 25: ...is provided with an interlock switch that will disconnect the illumination system immediately as soon as the upper part of the housing is removed 5 1 Removing the lamp module from the microscope and o...

Page 26: ...module Fig 5 3 1 into the mount up to the stop position here for example the mount for channel Tighten the knurled screw Fig 5 3 2 slightly to hold the module in place Connect the cable of the LED mo...

Page 27: ...Carl Zeiss Inserting or changing LED modules Colibri 2 26 423052 7244 001 07 2014 Fig 5 3 Inserting an LED module into the lamp module Artisan Technology Group Quality Instrumentation Guaranteed 888 8...

Page 28: ...ig 5 4 2 of the lamp module until it locks in place point of pressure Avoid tilting the mounting frame to prevent damage To change the mounting frame hold it by the grip and remove it carefully upward...

Page 29: ...ed or an already mounted beam combiner must be dismounted To insert or change the beam combiner proceeds as follows Please handle the beam combiner and the mounting frame respectively with reasonable...

Page 30: ...on the Colibri icon with the right mouse key MTB 2004 can also be used to configure the beam combiners and to register any additional excitation filters possibly mounted in front of the LED modules Fi...

Page 31: ...n selecting the corresponding beam combiner make sure that its wavelength is between the wavelengths of the adjacent LED modules for example beam combiner 425 nm between LED 365 nm on position 1 and L...

Page 32: ...Off 2 Power in 3 Overlaying the graphical representation 4 Switching off all LED modules 5 Selection between LED modules and external white light source 6 Selection between operating mode continuous...

Page 33: ...h 4 Selection between LED and external white light source 5 All LEDs On Off 6 External white light source activated 7 Illumination mode Gated activated Fig 6 4 Operating the LED modules in the illumin...

Page 34: ...is switched to the external white light source by clicking on External the LED modules will be switched off automatically The switching prism symbolized in the graphics changes its position to extern...

Page 35: ...mple configuration LED Pos 1 BC 12 LED Pos 2 BC 1234 LED Pos 3 BC 34 LED Pos 4 4 LED modules provided 1 365 nm 380 nm 400 nm BC 425 455 nm 470 nm BC 490 505 nm 530 nm BC 565 590 nm 615 nm 625 nm 3 LED...

Page 36: ...nm 400 nm BC 425 455 nm 470 nm 505 nm 530 nm 590 nm 615 nm 625 nm 6a empty empty 365 nm 380 nm 400 nm BC 425 empty empty 455 nm 470 nm 505 nm 530 nm 590 nm 615 nm 625 nm 7 empty empty empty empty 365...

Page 37: ...E Carl Zeiss Operation via the AxioVision Software Colibri 2 36 423052 7244 001 07 2014 Example 1 Example 2 Example 3 Example 4 Example 5 Example 6 Artisan Technology Group Quality Instrumentation Gua...

Page 38: ...9141 000 LED 420 nm 410 420 nm 410 420 nm 15 nm 423052 9131 000 LED 445 nm 450 nm 440 460 nm 24 nm 423052 9042 000 LED 455 nm 455 460 nm 450 455 nm 25 nm 423052 9052 000 LED 470 nm 470 475 nm 465 470...

Page 39: ...tter dual band 1 emission filter dual band 489060 9901 000 Filter set 60 HE CFP YFP HcRed shift free DMR 25 CFP YFP HcRed 3 excitation filters single band 1 beam splitter triple band 1 emission filter...

Page 40: ...imethyl carbinol 2 hydroxypropane To clean the optical surface moves the swab or cloth in circles from the center to the edge of the optical surface Slight pressure should be used on the optics during...

Page 41: ...ive fuse T 2 5 A H Push the fuse holder back into the fuse compartment until it locks into place Connect the power plug again 8 3 Service Any intervention concerning optical components or moving eleme...

Page 42: ...ces parts of devices or any procedures described in this manual are protected by the following patents US 6154282 A1 Semiconductor based excitation illuminator for fluorescence and phosphorescence mic...

Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 44: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Reviews: