![background image](http://html1.mh-extra.com/html/zeiss/colibri-2/colibri-2_operating-manual_3307572030.webp)
CONFIGURATION AND OPERATION IN MTB AND AXIOVISION
Colibri.2
Configuration of LED modules, beam combiners and excitation filters
Carl Zeiss
07/2014
423052-7244-001
29
6
CONFIGURATION AND OPERATION IN MTB AND AXIOVISION
6.1
Configuration of LED modules, beam combiners and excitation filters
The LED modules inserted into the lamp module are recognized and registered automatically by the
Auto
Configuration
function, provided that the control unit has been connected to the PC via the CAN bus of
the (motorized) stand. If the connection has been established directly to the PC via USB, Colibri has to be
selected manually in the MTB system and the LED modules have to be registered using the
Update
Configuration
function (clicking on the
Colibri
icon with the right mouse key).
MTB 2004 can also be used to configure the beam combiners and to register any additional excitation
filters possibly mounted in front of the LED modules.
Fig. 6-1
Configuration via MTB2004
Like other fluorescence illuminators, the Colibri.2 illumination system is used in the reflected light path
and appears, therefore, in the directory
Reflected Light
(Fig. 6-1/
1
).
•
By clicking on the
Colibri
icon (Fig. 6-1/
2
) you get access to the subordinated selection list with
various LED module and beam combiner positions.
•
When clicking on the submenu item
(No) Excitation Filter
(Fig. 6-1/
3
), a selection list of excitation
filters, which are suitable for being used in combination with the respective LED module, will appear in
the right field (Fig. 6-1/
4
). The filter actually used can be selected by a double click.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com