START-UP
Carl Zeiss
Attaching the lamp module, connecting the control unit and the control panel
Colibri.2
16
423052-7244-001
07/2014
3.2
Attaching the lamp module,
connecting the control unit and the
control panel
The lamp module operates via the illumination
mount (reflected light) of the microscope.
•
Attach the lamp module (Fig. 3-2/
1
) via the
dovetail of the illumination adapter to the
reflected light mount (Fig. 3-2/
4
) of the
microscope and fasten it by means of the
clamping screw.
The lamp module does not require any adjustment.
•
Connect the lamp module cable (Fig. 3-2/
2
) to
the 25-pin subD port of the control unit (see
Fig. 3-7/
4
).
•
Connect the connection cable of the control
panel (see Fig. 3-9/
1
) to the 15-pin subD port of
the control unit (see Fig. 3-6/
2
).
•
Connect the control unit (see Fig. 3-7/
2
) with
the power cable to a power outlet.
An external white light source (e.g. HXP 120 V, see Fig. 3-10) can be connected at the coupling position
for liquid light guides as follows:
•
Introduce one end of the light guide into the white light source (see Fig. 3-11/
3
) and the other end
into the connection point of the lamp module (Fig. 3-2/
3
) and fasten it by means of the swivel nut.
•
The internal motorized shutter of the white light source HXP 120 V can be triggered via the 9-pin
subD port of the lamp housing. Connect for this purpose the trigger cable with the BNC socket
OUT 1
of the Colibri control unit (see Fig. 3-7/
1
).
•
Connect the white light source (see Fig. 3-11/
2
) with a power cable to a power outlet.