4.5.4 BNC Trigger interface
The CL LED 9000 is fi tted with a BNC Trigger interface (14). The interface is located at the device rear and marked „Trigger“. This
interface allows the LED to be switched on and off quickly by means of an electrical signal (up to 5 V). For triggering e.g. Zeiss
Axiocam camera can be used. In addition, the BNC interface is compatible to the exposure function of the Zeiss Cell Observers.
For further information please contact Carl Zeiss.
4.6 Accessories
A comprehensive accessories programme for CL 6000 LED | CL 9000 LED is available from Carl Zeiss and authorised Zeiss
dealers. For more detailed information please have a look at our brochures about Zeiss stereo microscopy, e.g. stereo Discovery.
Flawless functional operation, safety and a high light yield of the light source are only ensured with original light guides recom-
mended by Carl Zeiss for the CL 6000 LED | CL 9000 LED.
CAUTION! Use of other light guides may trigger the safety shutdown which inhibits the switching-on of the il-
luminant of the cold light source. In such case the message „Light Guide!“ is shown in the display.
The original light guides are available from Carl Zeiss or authorised Zeiss dealers..
5 Maintenance
The CL 6000 LED | CL 9000 LED is basically a maintenance-free device.
The cold light source is not designed for desinfection (e.g. in medical applications). Please clean the cold light source only exter-
nally with a soft dry cloth.
Slightly moisten the cloth with a commercial plastic cleaning agent to remove tenacious soiling.
CAUTION! Do not spray any fl uids into the ventilation openings (7, 8). Do not use a wet cloth to clean the cold
light source.