With the footswitch (accessory) connected the light source can be switch on and off quickly and
easily. After switch-off of the light source by the footswitch, „Standby“ will be shown in the display.
(Also applicable to CL 9000 LED)
If no light guide was inserted into the light guide holder (4) or if the light guide was not inserted up
to the limit-stop or if a light guide of another manufacturer is used which is not suitable for the
CL 6000 LED | CL 9000 LED the light of the light source will possibly not light up upon switch-on,
and the display (3) will show the message „Light Guide!“.
Loosen the collet chuck by rotating it in counter-clockwise direction and insert a light guide into
the light guide holder (4) pushing it all the way to the limit-stop. Please note that only light guides
should be used which have been approved by Carl Zeiss for use with CL 6000 LED and CL 9000
LED. A list of these light guides can be found in the Zeiss brochures, system diagrammes and
price lists.
(Also applicable to CL 9000 LED)
4.3.7 Other messages shown in the display
CL 6000 LED
* Standby *
CL 6000 LED
Light Guide!
4.3.8 Error messages
The following error messages will help the service staff to identify a malfunction of the light guide and to rectify the problem quickly.
If one of the below error messages is shown in the display of the light guide, shutt off the device and contact customer service.
All error messages also apply to CL 9000 LED.
The integrated temperature sensor is not working properly, i.e. there is no protection against
potential overheating of the device.
The LED is defect or is disconnected from power supply.
The illuminant has heated up too much. The illuminant was switched off automatically. Make sure
the ventilating openings are not covered.
The control PCB has heated up too much. The illuminant was switched off automatically. Make
sure the ventilating openings are not covered.
The fan has become blocked or is malfunctioning. Suffi cient cooling is not ensured.
CL 6000 LED
Temp-Sensor!
CL 6000 LED
LED open!
CL 6000 LED
LED Overtemp
CL 6000 LED
PCB Overtemp
CL 6000 LED
Fan error!