2 Important notes
Symbols:
Intended use:
The cold light source CL 6000 LED | CL 9000 LED is intended for indoor use in industry and laboratories. The cold light source
must not be used in potentially explosive rooms or environments.
When used to the intended purpose, the cold light source serves to illuminate objects of all kinds by directing high-intensity visible
light through various light guides to the object.
The cold light source CL 6000 LED | CL 9000 LED meets the requirements of the following standards and directives:
DIN EN 61010-1 (IEC 61010-1)
DIN EN 62471
DIN EN 61326-1
Directive 2004/108/EC
Directive 2006/95/EC
Compliance of the device with requirements in the above-mentioned standards and directives is evidenced by means of the per-
tinent technical documentation.
Safety notes:
The cold light source emits visible high-intensity light. Metering of photo-biological safety according to DIN EN 62471 resulted in
the classifi cation of the device into Risk Group 2.
Please study and observe these Operating Instructions carefully, as they contain important information relating to the safety of the
user and the device. Safety of the device and the user is not ensured if the Operating Instructions are not adhered to.
CAUTION! Potentially hazardous optical radiation. Avoid looking into the lamp or the light guide aperture emis-
sion for longer periods during operation. May be harmful to the eyes.
Light-absorbing materials convert impinging light into heat. Hence, heat-sensitive and/or infl ammable light-absorbing materials
may suffer thermal damage and/or burns. Please observe the below notes in order to avoid such damage and hazards:
Symbol
Description
Warning of a hazard, observe documentation!