Auto Exposure 

  sets max pixels to 80% (not 100%) 

(can also be applied to a selected area)

 

Auto Contrast 

 unlike 

Gain 

only

 affects appearance of the image on the screen(!!!) 

Keypad controls: 

  top level button 1 = transmitted light on/off (#2-4 = empty) 

  FW button switches between control of: 

  FW1 = excitation filter wheel (orbit excitation filters, only used for FRET or CFP+YFP) 

  FW2 = emission filter wheel in front of camera, underneath  microscope 

Æ

 the orbit excitation filters must be in HOME position for use of the emission filters; if 

they aren’t, press the HOME button on the external keyboard

 

Control buttons on the lefthand side of the microscope: 

 

lowest button

  switches between frontport (empty) <> eyepiece <> baseport (camera) 

 

middle button

  switches between sideport 100% (empty) <> 50/50 <> eye/ front/ base 100% 

 

top:  

Optovar in / out 

File formats 

  The original Volocity libraries are stored as "mvd2" files together with a folder containing the 

image data in "aisf" files 

 TIFF

: 

looses all enhancements but keeps original ("raw") data 

  all stacks/ channels are stored in 1 multipage TIFF with the original bit-depth (8 or 12 bit) 

  TIFF for publication 

  screen version of image, including colours, merge, enhancements, on-screen image 

resolution, it is always 8-bit 

Æ

 loss of original ("raw") data!!!

 

merge colours: 

  CREATE NEW IMAGE SEQUENCE > MULTIPLE CHANNELS 

  drag single channel (or z / timepoint) images onto new image sequence window 

Summary of Contents for Axiovert 200

Page 1: ...are 3 General informations 3 Acquiring the image 3 refocussing re centring 3 adjusting channel presets 4 multidimensional acquisition setup 4 Shutdown procedure 5 Important additional information 7 Vo...

Page 2: ...on with the joystick holding the button on top of the joystick makes the stage moving faster focus on your sample if focus beeps override the top limit by pushing the top focus button on the right sid...

Page 3: ...ters filters focus except the light path BINOCULAR FRONT PORT BASE PORT and the objectives Acquiring the image start the VOLOCITY software make sure the dongle is attached to a USB port click on the c...

Page 4: ...open Acquisition Setup dialogue o tick Change channels o add as many channels as you need with the button o from the appearing drop down menus choose the desired channel presets o if you don t want to...

Page 5: ...tage menu in the Acquisition Setup dialogue set Change XY using to Ludl XY Stage o handmade positioning Video menu Live Video Stage menu Clear All Points move to desired point Stage menu Add Point wit...

Page 6: ...meone s booked within two hours update usage in Sharepoint remove your samples clean objective lenses with fresh lens tissue and close incubation chamber clear up the desk save files onto the server o...

Page 7: ...18 Feb 2011 qsg wf1 p 7 8 Important additional information Volocity Video Toolbar...

Page 8: ...rd Control buttons on the lefthand side of the microscope lowest button switches between frontport empty eyepiece baseport camera middle button switches between sideport 100 empty 50 50 eye front base...

Reviews: