Observing fluorescence through the eyepiece 

There are 2 options to observe fluorescence: the common way would be to use the filter cubes 
in the cube revolver; for some applications the use of a dichroic mirror in the filter cube position 
together with the filters in the filter wheels is necessary. The latter setup will only work with the 
camera, not with the eyepiece. 

  make sure the excitation / emission filter wheels are in the open position, 

unless needed: 

  on the filter wheel pad, select FW1 and then HOME 

  repeat for FW2 

  with the REFLECTOR buttons choose the illumination filter (microscope, 

right / bottom behind focus wheel) 

  switch between brightfield and fluorescence 

with FL ON / OFF (left / front / top button on 
microscope) 

 

Æ

 turn fluorescence immediately off if not 

needed, to avoid bleaching! 

  find your sample  

 

 

Acquiring images with the 
Improvision Volocity software 

General informations 

All 

 microscope parts can be controlled by the software (shutters, filters, focus, …), except the light 

path (BINOCULAR/ FRONT PORT / BASE PORT) and the objectives 

Acquiring the image 

  start the VOLOCITY software (make sure the dongle is attached to a USB port) 

  click on the camera icon to activate the live view and microscope controls(for more information 

about the Video Toolbar in the upcoming window see Additional Information section) 

  on the microscope stand, change the light path to BASEPORT 

  refocus/ re-centre (if needed, see "refoccussing/ re-centering") 

  adjust the channel presets of the channels you need (see "adjusting channel presets") 

  adjust other acquisition parameters (see "multidimensional acquisition setup") 

  with a click on the camera icon on the video toolbar you capture a single frame with the 

settings actually displayed in the video toolbar 

  with a click on the red button icon on the video toolbar you start your recording as defined in 

"Acquisition Setup" 

Useful: by clicking the "Freeze" button in the software you can keep your actual live image 
freezed on the screen; the acquisition stops and the shutters will close, preventing your 
sample from bleaching. 

refocussing/ re-centring: 

o

  choose the brightfield preset (preset button 1) 

Summary of Contents for Axiovert 200

Page 1: ...are 3 General informations 3 Acquiring the image 3 refocussing re centring 3 adjusting channel presets 4 multidimensional acquisition setup 4 Shutdown procedure 5 Important additional information 7 Vo...

Page 2: ...on with the joystick holding the button on top of the joystick makes the stage moving faster focus on your sample if focus beeps override the top limit by pushing the top focus button on the right sid...

Page 3: ...ters filters focus except the light path BINOCULAR FRONT PORT BASE PORT and the objectives Acquiring the image start the VOLOCITY software make sure the dongle is attached to a USB port click on the c...

Page 4: ...open Acquisition Setup dialogue o tick Change channels o add as many channels as you need with the button o from the appearing drop down menus choose the desired channel presets o if you don t want to...

Page 5: ...tage menu in the Acquisition Setup dialogue set Change XY using to Ludl XY Stage o handmade positioning Video menu Live Video Stage menu Clear All Points move to desired point Stage menu Add Point wit...

Page 6: ...meone s booked within two hours update usage in Sharepoint remove your samples clean objective lenses with fresh lens tissue and close incubation chamber clear up the desk save files onto the server o...

Page 7: ...18 Feb 2011 qsg wf1 p 7 8 Important additional information Volocity Video Toolbar...

Page 8: ...rd Control buttons on the lefthand side of the microscope lowest button switches between frontport empty eyepiece baseport camera middle button switches between sideport 100 empty 50 50 eye front base...

Reviews: