Startup procedure 

 

Turn on (in this order): 

 if 

required: 

heater unit

 and 

heater

, 4 hours before live imaging to allow focus to stabilise: 

 

c

 plate holder 

 

d

 chamber 

 if 

required: 

CO

2

 supply 

  switch on the mercury 

fluorescence lamp

 

  MAIN switch on the wall, on the right side of the microscope 

 computer 

Log in with your local account (not your IC login) and start the Volocity software. 

Adjusting brightfield Köhler illumination 

  make sure the LIGHT PATH (button on the left / bottom) is switched to 100% 

BINOCULAR: press button until displayed on the microscope top display 

  switch to 10x objective with the objective buttons on the right side of the 

microscope 

  you can move your specimen in xy-direction with the joystick, holding the 

button on top of the joystick makes the stage moving faster 

  focus on your sample 

 

if focus beeps, override the top limit by pushing the top focus button on the 
right side of the microscope while turning the focus wheel

 (when you hear double beeps, the 

mechanical limit of the focus is reached which cannot be overrun) 

  open brightfield shutter with S1 (labelled ‘BF’ on the filter wheel pad) 

  adjust brightness with large black button on front / bottom of 

microscope) 

  make sure condenser iris is not completely closed (two buttons on 

top of illumination filter wheel)  

  completely close field iris (top silver lever) (if the image turns 

completely black, reopen until you see some light, proceed to next 
step and close again after you have focussed the iris) 

  focus iris (black condenser focus wheel) 

  centre iris (2 silver Köhler screws) 

  reopen field iris until the edge is just not visible anymore 

Summary of Contents for Axiovert 200

Page 1: ...are 3 General informations 3 Acquiring the image 3 refocussing re centring 3 adjusting channel presets 4 multidimensional acquisition setup 4 Shutdown procedure 5 Important additional information 7 Vo...

Page 2: ...on with the joystick holding the button on top of the joystick makes the stage moving faster focus on your sample if focus beeps override the top limit by pushing the top focus button on the right sid...

Page 3: ...ters filters focus except the light path BINOCULAR FRONT PORT BASE PORT and the objectives Acquiring the image start the VOLOCITY software make sure the dongle is attached to a USB port click on the c...

Page 4: ...open Acquisition Setup dialogue o tick Change channels o add as many channels as you need with the button o from the appearing drop down menus choose the desired channel presets o if you don t want to...

Page 5: ...tage menu in the Acquisition Setup dialogue set Change XY using to Ludl XY Stage o handmade positioning Video menu Live Video Stage menu Clear All Points move to desired point Stage menu Add Point wit...

Page 6: ...meone s booked within two hours update usage in Sharepoint remove your samples clean objective lenses with fresh lens tissue and close incubation chamber clear up the desk save files onto the server o...

Page 7: ...18 Feb 2011 qsg wf1 p 7 8 Important additional information Volocity Video Toolbar...

Page 8: ...rd Control buttons on the lefthand side of the microscope lowest button switches between frontport empty eyepiece baseport camera middle button switches between sideport 100 empty 50 50 eye front base...

Reviews: