AxioCam  ERc 5s
Microscope camera

Operating instructions

Imprint

Carl Zeiss Microscopy GmbH 
Carl-Zeiss-Promenade 10 
07745 Jena, Germany

[email protected]
www.zeiss.de/axiocam 

Carl Zeiss Microscopy GmbH 
Königsallee 9-21 
37081 Göttingen, Germany

426540-7144-001
Printed 04/2012

© Carl Zeiss Microscopy GmbH, 2010

Control elements, indicators and con-

nections

Front of camera:
1  LED indicator
2  Manual acquisition
3   AWB:  Automatic  white/

gray balance

4   Receiver for IR remote 

control

5   Card slot for SD/SDHC 

memory cards

Back of camera:
1   USB port for data trans-

fer and power supply

2   DVI signal via HDMI 

port for monitor, TV 
or beamer

3   S-Video connection for 

recorder or monitor

Definitions

 
SD memory card:

 Digital storage medium for recording image data and 

other information. On SD cards an amount of 4 GB of data can be stored; up to 
32 GB on new SDHC cards (HC=high capacity). SD cards should be formatted on 
a PC to FAT32 format.

C-mount adapter: 

The C-mount is a standardized mounting thread for 

camera lenses. In microscopy the C-mount is used to connect digital cameras. 
So it is guaranteed that the image produced by the microscope objective is large 
enough to illuminate the small camera sensors.

AWB (Automatic White Balance):

 The purpose of the white balance is to 

calibrate the camera to the color temperature of the incident light and thereby 
prevent color cast. However, the automatic white balance that is normally 
used does not work, or does not work sufficiently well, in the case of technical 
surfaces or macros. This is because these objects often do not have any defined 
pure-white or color-neutral surfaces that the automatic white balance can take 
as a basis for the calibration. Instead it uses the average values of the individual 
red, green and blue (RGB) color channels for reference. It is therefore necessary 
to perform the white balance manually using defined color-neutral objects.

Parfocality:

 According to Professor A. Köhler, the focus plane set should be 

retained when eyepieces or objectives are changed on the microscope. Ideally, 
the object should appear just as sharp as it did before the eyepiece or objective 
was changed. Today, this also applies to the phototube for the acquisition of 
images using a digital camera: the sharpness of the camera’s images should 
correspond to that of the image you see when you look through the eyepiece. It 
is possible to set the parfocality of the camera even with the help of a reticle in 
the intermediate image of the eyepiece or with the help of an adapted monitor.

1

2

3

1

2

3

4

5

Infoflip is a registered trademark. This Infoflip is Made in Germany by Infoflip Medien GmbH. · IF.G.13.00617.E.02

Summary of Contents for AxioCam ERc 5s

Page 1: ...The purpose of the white balance is to calibrate the camera to the color temperature of the incident light and thereby prevent color cast However the automatic white balance that is normally used doe...

Page 2: ...ith the manufacturer concerned Always follow the instructions for the microscopes when it comes to connecting and operating external cameras Damage due to short circuit If the cable of the mains power...

Page 3: ...observation and recording devices by means of various standardized data interfaces Only pre fabricated standard connection cables may be used Obligations assurances and exclusions The envisaged uses o...

Page 4: ...Included with product Included with product Included in standard package AxioCam ERc5s camera with dust cap A to Mini B type USB cable 1 Camera 2 USB cable In addition DVD containing the configuration...

Page 5: ...hat may be present from the microscope s phototube 2 Place the unit comprising the camera and C mount adapter onto the microscope s phototube 3 Align the camera by rotating it so that the front of the...

Page 6: ...ve problems with the AWB please use a conversion filter to adjust the color temperature of the illumination to the camera limit Operation without PC connection Operation without PC connection Operatin...

Page 7: ...setup gives you the option of installing the configuration program for the camera the free imaging software AxioVision LE or both together To do so follow the instructions in the dialog boxes on the s...

Page 8: ...n the case of darkfield images 3 Saving and closing This button gives you the option of transferring the settings currently specified on the property pages to the camera s internal memory and then clo...

Page 9: ...opening so that it is impermeable to light Saving property page Naming Enter a prefix into the text field that will appear in front of the image number in the file name on the memory card Resolution...

Page 10: ...interface in US television format General Settings property page Reload Factory Defaults Resets camera to a defined status Note Your individual settings will be lost during this process Time And Date...

Page 11: ...e image subsequently click on the Properties button and then on the Display property page Adjust the image individually Brightness and Contrast sliders Set brightness and contrast automatically Min Ma...

Page 12: ...are update does not work An SD card must be inserted for an update to take place Insert a formatted unprotected SD card with at least 10 MB of free memory space No live image of the camera is visible...

Page 13: ...ol or maintenance elements on the camera housing or inside the camera that the user would be able to repair him herself If you are unable to resolve a problem using the troubleshooting table on the ba...

Page 14: ...For your notes For your notes For your notes...

Page 15: ...routine microscopy and image analysis The camera should only be installed and operated by trained individuals who have been informed about the possible risks associated with microscopy and the area o...

Reviews: