background image

Configuring and controlling the camera 

from the PC

> Connecting the power supply and establishing  

  a PC connection

To control the camera from a PC it must be connected to the PC by means of a USB 
cable. The PC also provides the power supply for the camera via this cable, meaning 
that there is no need for a separate power supply.

Caution!

Serious damage to the camera as a result of power spikes and electrostatics! When you 

connect the PC and camera, serious damage can be caused to the camera. In order to 

avoid this, please observe the following sequence when inserting the connectors!

 Switch off the PC in order to prevent damage caused by power spikes and 
electrostatics.

Check the USB cable for any signs of external damage.

Plug the flat USB connector into a free USB port on the PC.

 Insert the Mini USB connector into the small port on the bottom left of the 
back of the camera. Make sure that the connector is aligned correctly, the 
wider edge must be at the top!

Now switch the PC back on. 

If you later wish to disconnect the PC and camera again, proceed as follows (always 
hold the connector while pulling, not the cable!):

First disconnect the connector from the camera housing.

Then pull the connector out of the PC port.

> Installing configuration program and  

  AxioVision LE

A DVD with a setup program is supplied. The setup gives you the option of installing 

the configuration program for the camera, the free imaging software AxioVision LE 

or both together. 
To do so, follow the instructions in the dialog boxes on the screen.
When you perform the setup, the camera driver will also be installed at the same 
time. In addition, when AxioVision LE is installed a check will be performed to 
see whether an older version of the program already exists. If it does, this will be 
updated.
A manual and online help for AxioVision LE can also be found on the DVD.

> Configuring the camera

It is possible to configure the camera from the PC using the configuration program. 

This allows a range of processes that take place in the camera to be preset in detail. 

These settings are then saved onto the camera from the configuration program. 
With the help of the configuration program it is possible for the stand-alone 
operation of the AxioCam ERc 5s to be individually configured. This means that, 
using this configuration program, you can change various settings for the images 

that you save directly onto the SD card when the button on the camera is pressed. 
When the program is closed the parameters that have been changed can be 
transferred onto the camera as presets (user settings) and used automatically for 
your next images.
If the camera is used in stand-alone mode, these settings determine how the 
camera stores the photos onto the memory card, even without a PC connection.

  Operation with PC connection and 

 software 

  Operation with PC connection and 

 software 

i

!

!

Summary of Contents for AxioCam ERc 5s

Page 1: ...The purpose of the white balance is to calibrate the camera to the color temperature of the incident light and thereby prevent color cast However the automatic white balance that is normally used doe...

Page 2: ...ith the manufacturer concerned Always follow the instructions for the microscopes when it comes to connecting and operating external cameras Damage due to short circuit If the cable of the mains power...

Page 3: ...observation and recording devices by means of various standardized data interfaces Only pre fabricated standard connection cables may be used Obligations assurances and exclusions The envisaged uses o...

Page 4: ...Included with product Included with product Included in standard package AxioCam ERc5s camera with dust cap A to Mini B type USB cable 1 Camera 2 USB cable In addition DVD containing the configuration...

Page 5: ...hat may be present from the microscope s phototube 2 Place the unit comprising the camera and C mount adapter onto the microscope s phototube 3 Align the camera by rotating it so that the front of the...

Page 6: ...ve problems with the AWB please use a conversion filter to adjust the color temperature of the illumination to the camera limit Operation without PC connection Operation without PC connection Operatin...

Page 7: ...setup gives you the option of installing the configuration program for the camera the free imaging software AxioVision LE or both together To do so follow the instructions in the dialog boxes on the s...

Page 8: ...n the case of darkfield images 3 Saving and closing This button gives you the option of transferring the settings currently specified on the property pages to the camera s internal memory and then clo...

Page 9: ...opening so that it is impermeable to light Saving property page Naming Enter a prefix into the text field that will appear in front of the image number in the file name on the memory card Resolution...

Page 10: ...interface in US television format General Settings property page Reload Factory Defaults Resets camera to a defined status Note Your individual settings will be lost during this process Time And Date...

Page 11: ...e image subsequently click on the Properties button and then on the Display property page Adjust the image individually Brightness and Contrast sliders Set brightness and contrast automatically Min Ma...

Page 12: ...are update does not work An SD card must be inserted for an update to take place Insert a formatted unprotected SD card with at least 10 MB of free memory space No live image of the camera is visible...

Page 13: ...ol or maintenance elements on the camera housing or inside the camera that the user would be able to repair him herself If you are unable to resolve a problem using the troubleshooting table on the ba...

Page 14: ...For your notes For your notes For your notes...

Page 15: ...routine microscopy and image analysis The camera should only be installed and operated by trained individuals who have been informed about the possible risks associated with microscopy and the area o...