“Extras” property page

Sharpness

Emphasizes the object’s contrast edges. 
A slight enhancement of sharpness has 
been preset. This compensates for the 
lack of sharpness that occurs in all digital 
cameras due to the principles behind 
their operation. Excessive sharpness can, 
however, make the image look unnatural! 

Review Time

Defines the duration for which moving 

images or object movements are 

displayed. This varies between 1, 2, 5 or 
10 seconds.

Calibration

Improves the image quality of acquired images by correcting defective sensor pixels 

and defining the black reference. The camera’s calibration process takes several 

minutes.

Note:

When you perform this function, the camera should already have been switched on for at 

least fifteen minutes. The camera must see an image that is completely black. To achieve 

this, close the light path to the camera on the microscope. It is better to detach the 

camera from the microscope for this calibration process and to cover the sensor opening 

so that it is impermeable to light.

“Saving” property page

Naming

Enter a prefix into the text field that will 

appear in front of the image number in 

the file name on the memory card.

Resolution

Determines the image resolution, file 

size and quality with which the detail is 
reproduced.

Image Compression

The higher the quality, the more space the 
image takes up on the memory card.

“Live Image” property page

An external monitor can be connected to the AxioCam ERc 5s to enable co-
observation. This can either take place via the analog S-Video interface or the digital 
HDMI interface.
The HDMI signals can be displayed on a standard PC monitor as a DVI signal via the 
video adapter, which is available as an option.

Note:

It is not possible to display an HD signal (1080p/i or 720p/i) via the digital interface. 

Furthermore, only one of the interfaces (analog or digital) can be used. Co-observation 

in parallel to the PC is, however, possible and the camera can be controlled via the 

software. The settings in the configuration program also have a direct influence here on 

the display on the connected monitor.

  Configuration program (continued)

  Configuration program (continued)

i

!

!

i

!

!

Summary of Contents for AxioCam ERc 5s

Page 1: ...The purpose of the white balance is to calibrate the camera to the color temperature of the incident light and thereby prevent color cast However the automatic white balance that is normally used doe...

Page 2: ...ith the manufacturer concerned Always follow the instructions for the microscopes when it comes to connecting and operating external cameras Damage due to short circuit If the cable of the mains power...

Page 3: ...observation and recording devices by means of various standardized data interfaces Only pre fabricated standard connection cables may be used Obligations assurances and exclusions The envisaged uses o...

Page 4: ...Included with product Included with product Included in standard package AxioCam ERc5s camera with dust cap A to Mini B type USB cable 1 Camera 2 USB cable In addition DVD containing the configuration...

Page 5: ...hat may be present from the microscope s phototube 2 Place the unit comprising the camera and C mount adapter onto the microscope s phototube 3 Align the camera by rotating it so that the front of the...

Page 6: ...ve problems with the AWB please use a conversion filter to adjust the color temperature of the illumination to the camera limit Operation without PC connection Operation without PC connection Operatin...

Page 7: ...setup gives you the option of installing the configuration program for the camera the free imaging software AxioVision LE or both together To do so follow the instructions in the dialog boxes on the s...

Page 8: ...n the case of darkfield images 3 Saving and closing This button gives you the option of transferring the settings currently specified on the property pages to the camera s internal memory and then clo...

Page 9: ...opening so that it is impermeable to light Saving property page Naming Enter a prefix into the text field that will appear in front of the image number in the file name on the memory card Resolution...

Page 10: ...interface in US television format General Settings property page Reload Factory Defaults Resets camera to a defined status Note Your individual settings will be lost during this process Time And Date...

Page 11: ...e image subsequently click on the Properties button and then on the Display property page Adjust the image individually Brightness and Contrast sliders Set brightness and contrast automatically Min Ma...

Page 12: ...are update does not work An SD card must be inserted for an update to take place Insert a formatted unprotected SD card with at least 10 MB of free memory space No live image of the camera is visible...

Page 13: ...ol or maintenance elements on the camera housing or inside the camera that the user would be able to repair him herself If you are unable to resolve a problem using the troubleshooting table on the ba...

Page 14: ...For your notes For your notes For your notes...

Page 15: ...routine microscopy and image analysis The camera should only be installed and operated by trained individuals who have been informed about the possible risks associated with microscopy and the area o...

Reviews: