DynaPak 3010RNM Version 01102001 ATEX Rev.
YZ Systems, Inc. • 3101 Pollok Drive • Conroe, Texas • USA • 77303 • P: 936.788.5593 • F: 936.788.5720
To set this value turn on switch positions
S2-2, S1-8, S1-7, S1-6, S1-5, S1-4, and S1-3.
Turning a switch "ON" enables the divisor value.
Switch divisor values are cumulative.
2300
-2048 S2-2 “ON”
252
- 128 S1-8 “ON”
124
- 64 S1-7 “ON”
60
- 32 S1-6 “ON”
28
- 16 S1-5 “ON”
12
- 8 S1-4 “ON”
4
- 4 S1-3 “ON”
0
The input signal threshold level is set from the
factory to work with most applications. However,
due to varying cable lengths and wire types, the
threshold level may need to be adjusted for
certain applications. The threshold level can be
adjusted using the “Input Adj.” potentiometer. If
pulses occur too often, not at all, and/or erratically,
you may need to change the threshold setting.
Turning the Input Adj. Potentiometer clockwise will
decrease the input sensitivity, and
counterclockwise will increase the input
sensitivity.
8. Sampler set-up:
Continuous proportional-to-flow sampling