Dissolved Oxygen
For best results, we recommend that the KCl solution and the membrane cap be
changed at least once every 30 days.
•
The rate of stirring should be at least 1 foot per second.
•
Membrane life depends on usage. Erratic readings are a result of loose, wrinkled,
damaged, or fouled membranes, or from large (more than 1/8" diameter) bubbles in
the electrolyte reservoir. The average replacement interval is two to four weeks.
•
If the membrane is coated with oxygen consuming (e.g. bacteria) or oxygen
producing organisms (e.g. algae), erroneous readings may occur.
•
Chlorine, sulfur dioxide, nitric oxide, and nitrous oxide can affect readings by
behaving like oxygen at the sensor. If you suspect erroneous readings, it may be
necessary to determine if these gases are the cause.
•
Avoid any environment that contains substances that may attack the probe module
and sensor materials. Some of these substances are concentrated acids, caustics,
and strong solvents. The sensor materials that come in contact with the sample
include FEP Teflon, acrylic plastic, EPR rubber, stainless steel, epoxy,
polyetherimide and the PVC cable covering.
Summary of Contents for YSI 556 MPS
Page 1: ...YSI 556 MPS Set up Operation Calibration...
Page 5: ...YSI 556 MPS SET UP...
Page 13: ...YSI 556 MPS SENSORS...
Page 19: ...View all Parameters...
Page 20: ...YSI 556 MPS CALIBRATION...
Page 35: ...YSI 556 MPS RUN...
Page 37: ...YSI 556 MPS FILE...
Page 45: ...YSI 556 MPS LOGGING...
Page 52: ...YSI 556 MPS SYSTEM SET UP...
Page 55: ...YSI 556 MPS MAINTENANCE...
Page 64: ...YSI 556 MPS STORAGE...
Page 68: ...Please visit the YSI website on WWW YSI COM Thank you...