Dissolved Oxygen
•
NOTE:
The instrument must be on for at least
20 minutes to polarize the DO sensor before
calibrating.
•
Use the arrow keys to highlight the
Dissolved
Oxygen
selection.
•
Press
Enter
. The dissolved oxygen calibration
screen is displayed.
•
Use the arrow keys to highlight the DO%
selection.
•
Press
Enter
. The DO Barometric Pressure
Entry Screen is displayed.
•
Place approximately 3 mm (1/8 inch) of water in
the bottom of the transport/calibration cup.
•
Place the probe module into the
transport/calibration cup.