Conductivity
•
Observe the reading under Specific
Conductance. When the reading shows no
significant change for approximately 30
seconds, press
Enter
. The screen will
indicate that the calibration has been
accepted and prompt you to press
Enter
again to Continue.
•
Press
Enter
. This returns you to the
Conductivity Calibrate Selection Screen.
•
Press
Escape
to return to the calibrate
menu.
•
Rinse the probe module and sensors in tap
or purified water and dry.