Conductivity
•
Use the arrow keys to highlight the
Conductivity
selection.
•
Press
Enter.
The Conductivity Calibration
Selection Screen is displayed.
•
Use the arrow keys to highlight the Specific
Conductance selection.
•
Press
Enter.
The Conductivity Calibration
Entry Screen is displayed.
•
Place the correct amount of conductivity
standard into a clean, dry or pre-rinsed
transport /calibration cup.
•
Carefully immerse the sensor end of the probe
module into the solution.
•
Gently rotate and/or move the probe module up
and down to remove any bubbles from the
conductivity cell.