pH
•
Use the arrow keys to highlight the
pH
selection.
•
Press
Enter
. The pH calibration screen is
displayed.
– Select the
1-point
option only if you are
adjusting a previous calibration.
– Select the
2-point
option to calibrate the pH
sensor using only two calibration standards.
Use this option if the media being monitored
is known to be either basic or acidic.
– Select the
3-point
option to calibrate the pH
sensor using three calibration solutions. In
this procedure, the pH sensor is calibrated
with a pH 7 buffer and two additional buffers.
The 3-point calibration method assures
maximum accuracy when the pH of the
media to be monitored cannot be
anticipated.