5.3 Changing the infusion set
The YpsoPump
®
Orbit
®
soft infusion set should not be used for more than 72 hours.
The YpsoPump
®
Orbit
®
micro infusion set should not be used for more than 48 hours.
When changing the infusion set, the insulin pump can either be in stop mode or in run mode.
Disconnect the infusion set as described in the instructions on page 93 and carefully remove the
tape from your body.
Then follow the steps on page 95.
Before attaching the tubing to your body, please make sure the tubing is filled with insulin without any
air bubbles.
Dispose of the introducer needle (YpsoPump
®
Orbit
®
soft), the used cannula base and the used tubing with adapter
according to the instructions given by your doctor or diabetes advisor and in compliance with the local regulations
for disposing of needles and sharp or pointed objects.
Only use the YpsoPump
®
Orbit
®
soft and YpsoPump
®
Orbit
®
micro infusion sets for therapy with the
YpsoPump
®
. All tubing and cannula lengths of these products are compatible with the YpsoPump
®
.