5. Changing the infusion set and cartridge
Warnings
Only rapid-acting insulin at a concentration of 100 U/ml (NovoRapid
®
, Humalog
®
, Apidra
®
) may be used with the
YpsoPump
®
system. Using different insulin can lead to incorrect dosages of insulin, which can cause hypoglycemia or
hyperglycemia (read warning in the box on page 7). If you use different insulin, change the cartridge as well as the
primed infusion set and continue therapy with the correct type of insulin.
With the YpsoPump
®
you may only use the self-filled 1.6-ml reservoir YpsoPump
®
Reservoir that is compatible with the
YpsoPump
®
. Using a different cartridge which is not declared by the manufacturer to be compatible with YpsoPump
®
can lead to incorrect dosages of insulin, which can cause hypoglycemia or hyperglycemia (read warning in the box on
page 7). In this case, change the cartridge.
Never use tools or instruments to remove the cartridge or the infusion set. They could damage the drive or the housing
of the YpsoPump
®
and compromise its operability. If the insulin pump is damaged, the correct delivery of insulin cannot
be guaranteed; this can cause hypoglycemia or hyperglycemia (read warning in the box on page 7). Ensure that
your delivery of insulin is as recommended by your doctor or diabetes advisor and consult Customer Service. You will
find the contact details on the back of this User Guide as well as on page 10.
Ensure the use of the correct drug solution and check the label of the vial before filling the YpsoPump
®
Reservoir.