![Ypsomed YpsoPump User Manual Download Page 67](http://html.mh-extra.com/html/ypsomed/ypsopump/ypsopump_user-manual_915220067.webp)
67
All bolus types functions are not available if the insulin pump is in stop mode, if the cartridge is empty, if the
remaining insulin amount in the cartridge is smaller than the bolus increment setting or if the bolus limit
setting is lower than the bolus increment setting. Bear in mind that entering a minimum bolus amount is directly
dependent on the bolus increment setting. If the bolus increment is set to 1.0 U, for example, the minimum
bolus amount that can be set for a standard bolus is also 1.0 U.
Standard bolus
2/4:
Tap the “Standard
bolus” icon.
1/4:
Open the main menu and tap the
“Bolus” icon.
3/4:
Select a bolus amount between
0.1 U and 30.0 U. Bolus delivery
commences immediately as soon as
you confirm by tapping . If the
default value of 0.0 U is confirmed,
the standard bolus is canceled.
The YpsoPump
®
vibrates briefly.