At night it may not be possible to perceive alarms if the insulin pump is not positioned properly (e.g. if you are lying on
the pump). Make sure the insulin pump is positioned properly while you are asleep so that you perceive any alarms.
Alarms that have not been dealt with can lead to failure of the preset insulin delivery, which can cause hyperglycemia
(read warning in the box on page 7).
Always make sure the infusion set is immediately disconnected from your body after an occlusion alarm in order to
prevent insulin from being delivered to the body when the occlusion is suddenly eliminated. Not disconnecting the
infusion set can lead to incorrect dosages of insulin, which can cause hypoglycemia or hyperglycemia (read warning in
the box on page 7).