1
2
4
3
35
3/7:
If you wish to reduce the current
basal rate, select a value between 0 %
and 90 %. If you wish to increase the
current basal rate, select a value between
110 % and 200 %, e.g. if you choose a
temporary basal rate of 80 %, the current
basal rate is reduced by 20 %. 100 %
means no reduction and no increase.
Then confirm by tapping .
4/7:
Select the period of time in which
you want to reduce or increase the current
basal rate. The period can be set
from 15 minutes to 24 hours in steps of
15 minutes.
How to activate a temporary basal rate
The temporary basal rate function allows to reduce or increase the basal rate for a
limited period of time. When the temporary basal rate set has finished, the programmed
basal rate profile (A or B) continues.
The temporary basal rate function is not available in stop mode. To use this function,
the insulin pump must be in run mode (see page 50).
1/7
Open the main menu and tap the
“Basal rates” icon.
2/7:
Tap the “Temporary basal rate
function” icon.