How to change the battery
2/5:
Open the battery compartment by
inserting a coin in the groove of the
battery compartment lid and turn
counterclockwise. Remove the empty
battery from the battery compartment
and dispose of it in accordance with your
national environmental protection
regulations.
1/5:
If the “Battery charge level low”
warning appears, the insulin pump can
still be operated for at least two days.
Replace the battery as soon as possible.
If the battery compartment has remained open for more than 5 minutes and the
mylife
™
YpsoPump
®
is in run mode, an alarm is triggered. After entering a new bat-
tery, all insulin pump settings are saved and do not have to be entered again.
If the battery compartment remains open for more than 5 minutes and the mylife
™
YpsoPump
®
is in stop mode, the insulin pump is automatically put into storage
status. When changing the alkaline battery, make sure you do not lose the battery
compartment lid.