66
Electronic error
If the mylife
™
YpsoPump
®
detects an internal malfunction, an electronic error is displayed.
All the functions of the insulin pump are cancelled. In the event of an electronic
error, the insulin pump displays two different screens alternately which show you how
to restart the insulin pump in a few steps.
Disconnect the infusion set from the
body and remove the alkaline battery
from the battery compartment.
Then press the function button for
2 sec onds. The insulin pump is now devoid
of a battery and in storage status. To
put the insulin pump back into operation
reinsert the alkaline battery, perform
the self-test and set the time of day and
the date. Check the therapy settings.
Change the cartridge and the infusion set
in accordance with the procedure on
page 37 of this Caregiver Guide.
If the electronic error remains after the insulin pump has been put into operation
again or if it reoccurs after a certain time, stop using the pump, remove the alkaline
battery and contact Customer Service. You find the contact details on the back of
this Caregiver Guide.