38
Do not insert the new cartridge until the threaded rod has completely returned and
the mylife
™
YpsoPump
®
has successfully performed the self-test. If you insert the
cartridge while the threaded rod is still returning, the insulin pump may display the
message “Threaded rod return not completed”. In this case repeat the cartridge
change.
8/8:
Remove the empty cartridge and the
infusion set from the mylife
™
YpsoPump
®
and dispose of them in accordance with
your national environmental protection
regulations.
7/8:
Disconnect the infusion set by
turning the adapter on the mylife
™
YpsoPump
®
counterclockwise until it
stops.
Inserting the cartridge and replacing the infusion set
1/2:
Hold the mylife
™
YpsoPump
®
upright
with the opening of the cartridge
compartment facing upwards. Insert a
self-filled 1.6-ml reservoir or pre-filled
1.6-ml insulin cartridge.
2/2:
Place the new adapter on the
inserted cartridge upright. Turn the
adapter clockwise to the lock position.
You should hear a soft click and/or
feel a definite mechanical stop.
1
7
2
8