YPP6830 Pulse Input Rate/Totalizer
Instruction Manual
39
K-Factor Value (
factr
)
Enter the K-factor value. This value is entered in Pulses/Unit as defined by the
K-
Factor Units
parameter. Most flowmeter manufacturers provide this information with
the device.
Display Units (
Units
)
The
Units
menu is used to select the display rate units and time (example: Gal/s) and
the display units for total and grand total.
Important Programming Note:
The units selected in this menu are the desired display units only. The units de-
fined by the k-factor of a flow meter are entered in the K-Factor menu as part of
the Factor Unit menu programming. See K-Factor Units (
FuNiT
) on page 38 for
details.
This allows the display units to be different than the units defined by the flow me-
ter, or be changed easily after initial programming. Unit conversions for rates and
totals are performed automatically by the meter. See Automatic Unit Conversions
on page 45 for details.