<2. WIRING AND INSTALLATION>
2-6
IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00
2.5 Wiring
Wire in the following steps. (Figure 1.3) (Each parenthesis shows the section number to see.)
1:
Power supply terminals (See 2.5.1)
2:
Relay module (2.5.2)
3:
Communication module (2.5.3)
4:
2nd sensor module (2.6)
5:
1st sensor module (2.6)
6:
IO module (2.5.4)
1st and 2nd order shown above are reversible. When wiring be sure to remove the shield cover.
After the wiring, screw up the shield cover.
WARNING
Use wiring cables with heat resistance above 75 °C, waterproof rating IP66 or higher.
Four front panel screws should be tightened to the following torque; 1.5 to 1.6 N·m.
For power supply, wire cables with a flammability rating of UL 2556 VW-1 or equivalent.
WARNING
Always place the shield cover over the power supply and contact terminals for interference
avoidance and safety reasons.
CAUTION
Turn off power supply to FLXA402 before wiring cables.
CAUTION
After wiring of power supply or contact output is completed, put on the shield cover. (see section
2.3)