<2. WIRING AND INSTALLATION>
2-2
IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00
2.2
Mounting methods
See 1.5 Dimensions for information on the hardware for mounting.
The FLXA402 has various mounting possibilities.(Figure.2.1)
• Panel mounting using optional mounting hardware (/PM or /UM)
• Wall mounting using optional mounting hardware (/U or /UM)
• Pipe mounting using optional mounting hardware (/U or /UM) on a horizontal or vertical
pipes (Nominal size: 50A)
The universal mounting kit (/UM) contains the pipe and wall mounting hardware (/U) and the
panel mounting hardware (/PM). Some can be left unused.
195
0
+1
138
0
+1
138
185
Unit: mm
Panel mounting
Wall mounting
Pipe mounting
Note: The attached hardware is the same
as that for the pipe mounting.
Only the bracket is used for the
wall mounting.
Note: The wall on which the
analyzer is mounted
should be strong enough
to bear the weight of
more than 8kg.
Note: Horizontal pipe mounting is
possible by turning the bracket
90 degrees.
M8 bolt (not supplied)
Prepare bolts of the same
length as the mounting hole.
Note: For a sun shade hood (option),
use the upper two bracket
mounting screws to fix it.
Bracket
Wall mounting holes (x3)
Wall
Bracket mounting screws*
Bracket mounting screws*
Fixing screws (x2)
Panel
Spacing panel cutout
Nuts (x2)
Washers (x2)
Stanchion (50A pipe)
Bracket mounting screws*
Bracket
Bracket
Bracket for piping
U-bolt
* Tighten the four screws to a torque of 2 N·m.
* Tighten the four screws to a torque of 2 N·m.
* Tighten the four screws to a torque of 2 N·m.
Figure 2.1
Three types of mounting with the optional mounting hardware
When you change the mounting type, you can order mounting hardware optionally. For further
information see
Summary of Contents for FLXA402