<2. WIRING AND INSTALLATION>
2-4
IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00
2.4
Installing the cable glands
The supplied cable glands are for cables with an outside diameter of 6 to 12 mm (0.24 to 0.47
inches). Unused cable entry holes must be sealed with cable glands including the supplied
rubber close up plugs.
When the suffix code for Digital Communication code is -E (Modbus TCP/IP), apply cable glands
(black) for Ethernet to the holes on the modules for communication.
For installing conduit adapters, apply Conduit adapter set (need to be purchased additionally,
seel Conduit adapter (need to be purchased additionally)
purchased additionally) and follow the instruction shown on Figure 2.5.
For IO module
For sensor modules
For relay module
For power supply
For COM module
Figure 2.3
Cable gland configuration
CAUTION
Be careful not to be injured by the sharp hole edges on the housing.
Cable gland
O-ring
Cable gland
cap
Case
Close up plug
(rubber)
Seal packing
Figure 2.4
Cable glands
CAUTION
To install cable glands, tighten it with torque of 2 N•m.