<2. WIRING AND INSTALLATION>
2-22
IM 12A01F01-02EN 2nd Edition : Feb. 19, 2019-00
2.6.4
Wiring the dissolved oxygen (DO) sensor
The input module for DO measurement is suitable for different types of DO sensors:
i. Galvanic sensors like model DO30G
ii. Polarographic sensors like HAMILTON’S Oxyferm and Oxygold
The connection is as follows:
11 TC +
TC -
+ anode galvanic
- cathode galvanic
+ anode polarographic
shield
- cathode polarographic
12
16
13
18
14
17
15
FLXA402
The DO30G sensor comes with integral cable and the wires are labeled with the corresponding
terminal numbers.