4.9 Burst Setting and Manipulation
119
IM FG410-01EN
c) Triggered gate oscillation setting screen
■
1st page: Screen for the basic parameters
These are the common setting items that are common to all the oscillation modes.
Oscillation start/stop phase [Phase]
This is the phase at which oscillation starts and stops.
■
2nd page: Screen for the triggered gate oscillation settings
Stop level [StpLvl]
This is the signal level during oscillation stop. The level can be set to either off or on.
Normally, it is set to off.
Trigger [Trig]
This is the trigger condition.
Internal and external can be chosen as the trigger source.
Oscillation stop unit [OscStop]
This is the oscillation stop unit. A choice can be made between 1-cycle unit and 0.5-cycle
units. Normally, 1-cycle unit is set.
Sync output [SyncOut]
This is the output signal from the sync/sub-output terminal. Waveform reference phase
synchronization and burst synchronization can be chosen from.
d) Waveforms for which triggered gate oscillation is not possible
Gate oscillation is not possible for DC.
Gate oscillation can be done for noise, but the operation differs from that of other
waveforms.
e) Setting items required for triggered gate oscillation
Set [Phase] (oscillation start/stop phase) on the 1st page of the setting screen.
[StepLvl] (stop level) on the 2nd page of the setting screen is set to [Off] normally.
[OscStop] (oscillation stop unit) on the 2nd page of the setting screen is set to [Cycle] (1
Set the oscillation
mode to [Burst]
Press the NEXT
key to display
the 2nd page
Oscillation start/stop
phase setting
Indicates that the 1st page
is displayed
Press the NEXT
key to display
the 1st page
Indicates that the 2nd
page is displayed
Set burst mode to
[TrigGate]
Stop level
Sync output
Oscillation stop unit
Trigger