106
4.
BASIC OPERATION
IM FG410-01EN
4.9.2 Auto burst
Oscillation and stop are automatically repeated with the respectively specified wave
number. No trigger signal is required.
Settings and manipulations are done on the Oscillator setting screen. “Oscillator” is
displayed at the top left of the Oscillator setting screen. When another screen is displayed,
pressing the
MENU key displays the top menu, so select [Oscillator] and then press
the
ENTER key.
a) Auto burst example
The following example is for mark wave number (oscillation wave number) = 3, space wave
number (oscillation stop wave number) = 2, oscillation start/stop phase = 30°, stop level =
off.
b) Setting the oscillation mode to auto burst
Set [Mode] (oscillation mode) to [Burst] on
the Oscillator setting screen. The oscilla-
tion mode is switched to the burst
oscillation mode as a result.
In the burst oscillation mode, the setting
screen consists of two pages, which can
be toggled with the
NEXT key.
Next, set [BrstMode] (burst mode) to
[Auto] (auto) on the 2nd page of the burst
oscillation mode.
MENU
ENTER
Oscillation start/stop phase
30°
Mark wave
number: 3
Oscillation
Space wave
number: 2
Stop
Oscillation phase continues even during stop
Select [Burst] in
[Mode] and then
press the ENTER
key
NEXT
Select [Auto] in
[BrstMode] and
then press the
ENTER key