IM 701310-02E
Working with the DL9000
Shifting the Trigger Position to the Left by Two Divisions
User’s manual section 6.2, “Setting the Trigger Position”
By shifting the waveform to the left by just 2 div, the part of the waveform after the trigger (post trigger) becomes
much easier to see.
1
Press the POSITION key.
Trigger position mark
Pre-trigger
portion
Post-trigger portion
Trigger point
2
Turn the rotary knob counterclockwise to
set Position to 30%.
POSITION
DELAY
Changing the Trigger Mode from Auto to Single
User’s manual section 6.1, “Setting the Trigger Mode”
When a trigger activates in Single mode, the waveform display update stops once, and waveform acquisition stops.
Single mode is useful for observing single-shot waveforms.
2
Select Single.
1
Press the TRIG
MODE/HOLD OFF key.
When waveform acquisition stops, the status changes from “Running” to “Stopped.”
TRIG MODE
HOLD OFF
ACQ COUNT ACTION